The advanced valve design of the ASI Self Priming Cartridge
Check Valve eliminates the need to manually prime HPLC pumps
with a syringe, even if the pump head and intake line are
completely dry. Large air bubbles in the intake line do not
cause loss of prime, and can be pumped to waste by opening the
purge valve. The valve features a crush proof design with a
15,000 psi maximum operating pressure rating. The valve can be
produced in variety of sizes to suit OEM applications.
Biocompatible and CO2 compatible valves are also available.